Keithley Instruments 2700 ROM Upgrade Software vB05 Notice for Model 7710 Support 11/18/2002 This file contains additional information regarding the installation and use of the ROM upgrade software for the Model 2700 Multimeter/ Data Acquisition System. Please review the contents of this file before installing or using the ROM Upgrade software. Introduction The Model 2700 mainframe requires firmware revision B05 or higher (B05, B06, etc.) to properly support the Model 7710 20-Channel Solid State/Long-Life Differential Multiplexer w/Automatic CJC switch module. If the firmware revision of your instrument is lower than B05 a ROM upgrade is required. This upgrade can also be used for those who wish to have the latest firmware enhancements and upgrades installed on their instrument. Step 1. Determine the firmware revision of your Model 2700 The firmware revision is briefly displayed during the power-on sequence of the instrument. When the instrument is turned on, all display annunciators will turn on for a short period of time. After that, the firmware revision will be briefly displayed as follows: REV: yyy zzz where: yyy is the mainframe revision (e.g., B04 of the Model 2700). zzz is the display board firmware revision (e.g., A01), which does not apply to this upgrade. To determine the firmware revision remotely, use the *IDN? query command. After sending the *IDN? command and addressing the instrument to talk, the following response message is sent to the computer: KEITHLEY INSTRUMENTS INC.,MODEL wwww,xxxxxxx,yyy/zzz where: wwww is the mainframe model (2700). xxxxxxx is the serial number of the mainframe. yyy is the firmware revision of the mainframe. zzz is the display board firmware revision. Step 2. Determine the next course of action Based on the firmware revision of your Model 2700, choose the appropriate course of action: a) If the firmware revision is B05 or higher on the Model 2700, an upgrade is not needed to use the Model 7710. b) If the firmware revision is B04 or lower on the Model 2700, an upgrade IS required. This upgrade can also be used for those who wish to have the latest firmware enhancements and upgrades installed on their instrument. Proceed to Step 3 to perform the upgrade. NOTE: If the firmware revision begins with the letter A (e.g., A01, A02, etc.) for the Model 2700, DO NOT perform the upgrade. For these revisions, a hardware upgrade is also required. Contact your local Keithley service center for instructions and pricing for upgrading your Model 2700. Step 3. Perform the ROM upgrade System requirements The minimum requirements include a Pentium-class PC running Windows 95 or later. The firmware can be installed via the IEEE-488 bus or the RS-232 interface: IEEE-488 Bus - If using the IEEE488 bus, a Keithley, CEC, or National Instruments IEEE-488.2 interface board must be installed in the PC. Use an IEEE488.2 cable to connect the interface board to the Model 2700 mainframe. RS-232 interface - If using the RS-232 port, use an RS-232C cable to connect the communications port to the Model 2700. NOTE: With RS-232, upgrade takes approximately 20 minutes. ROM upgrade procedure 1. From Windows, click Start->Settings->Control Panel>Add/ Remove Programs, and then remove any Model 2700 ROM upgrades if previously installed. NOTE: All saved scanlists, setups, and buffer readings will be lost with this upgrade. 2. Download the 2700 B05 Firmware upgrade.exe file from the Keithley Web site, www.keithley.com to a temporary directory on your PC. 3. From Windows click Start->Run. Click the Browse button and navigate to temporary directory and select the 2700 B05 Firmware upgrade.exe file. This should return you to the Run dialog box. Click OK to run the WinZip Self-Extractor. Specify a temporary directory to unzip the install files to and click Unzip. When the unzip operation has finished click OK and then Close. 4. From Windows click Start->Run. Click the Browse button and navigate to the temporary directory where you unzipped the files and select the setup.exe file. This should return you to the Run dialog box. Click OK to run the setup and follow all prompted instructions. 5. From Windows click Start->Programs->Keithley Instruments-> Model 2700 Flash Upgrade vB05 and follow all prompted instructions. 6. Perform Step 1 above (determine the firmware revision of your mainframe) to verify correct installation. NOTE: Error +516,"Battery Backed Memory Lost" will be displayed after the first power cycle after upgrade. Known issues After upgrading using the RS-232 interface the ROM upgrade software may report a 'Timeout Error - Error In Programming Model 2700' error. This error can be ignored. This is only a cosmetic error and has no effect on the successful completion of the upgrade procedure. This issue has been resolved and will be released in future firmware revisions.